A comparison of photometric methods for serum iron determination under flow analysis conditions

Autor
Rybkowska-Kucharczyk, Natalia
Strzelak, Kamil
Koncki, Robert
Data publikacji
2018
Abstrakt (EN)

In this contribution, a multicommutated flow analysis (MCFA) system for serum iron determination is presented. The construction of the system has enabled sequential measurements with three different determination methods with 1,10-phenantroline, Ferrozine and Ferene S using separate analytical units (modules). For each method a dedicated optoelectronic detector has been developed. Under optimal flow conditions, investigated methods were analytically characterized by linear range and sensitivity: 1.5 − 100.0 μmol L−1 and 4.9 mV L μmol−1, 3.3 − 50.0 μmol L−1 and 6.6 mV L μmol−1, 4.0 − 40.0 μmol L−1 and 13.3 mV L μmol−1 for 1,10-phenantroline, ferrozine and ferene S methods, respectively. In the course of investigations also clinical usefulness of each method has been discussed. The analysis of serum samples proved that Ferrozine and Ferene S methods are sufficient for serum iron determination with relatively low consumption of samples (less than 50 μL of serum).

Słowa kluczowe EN
Serum iron
Multicommutation
Flow analysis
Optoelectronic detector
Dyscyplina PBN
nauki chemiczne