Artykuł w czasopiśmie
Brak miniatury
Licencja

CC-BYCC-BY - Uznanie autorstwa

Unintended Changes of Ion-Selective Membranes Composition—Origin and Effect on Analytical Performance

Autor
Maksymiuk, Krzysztof
Stelmach, Emilia
Michalska-Maksymiuk, Agata
Data publikacji
2020
Abstrakt (EN)

Ion-selective membranes, as used in potentiometric sensors, are mixtures of a few important constituents in a carefully balanced proportion. The changes of composition of the ion-selective membrane, both qualitative and quantitative, aect the analytical performance of sensors. Dierent constructions and materials applied to improve sensors result in specific conditions of membrane formation, in consequence, potentially can result in uncontrolled modification of the membrane composition. Clearly, these eects need to be considered, especially if preparation of miniaturized, potentially disposable internal-solution free sensors is considered. Furthermore, membrane composition changes can occur during the normal operation of sensors—accumulation of species as well as release need to be taken into account, regardless of the construction of sensors used. Issues related to spontaneous changes of membrane composition that can occur during sensor construction, pre-treatment and their operation, seem to be underestimated in the subject literature. The aim of this work is to summarize available data related to potentiometric sensors and highlight the eects that can potentially be important also for other sensors using ion-selective membranes, e.g., optodes or voltammetric sensors.

Słowa kluczowe EN
ion-selective membranes
components leakage
incorporation
all-solid-state sensors
Dyscyplina PBN
nauki chemiczne
Czasopismo
Membranes
Tom
10
Zeszyt
10
Strony od-do
266
ISSN
2077-0375
Data udostępnienia w otwartym dostępie
2020-09-28
Licencja otwartego dostępu
Uznanie autorstwa